Published by Applied Physics Corporation · Manufacturer since 1992Technical sales: 719-428-4042 · [email protected]
Barrier systems

Isolator and RABS Airflow Visualization

Isolators and RABS concentrate the airflow question into a smaller, higher-consequence space. The study must show first air, operator interaction, transfer routes and the effect of openings without letting the fog-introduction method dominate the enclosure.

Map the critical pathways

  • HEPA source to exposed product or critical surface
  • Glove and sleeve movements through first air
  • Door, mouse-hole and transfer-port openings
  • Equipment and container blockage
  • Interventions, replenishment and removals
  • Return or exhaust locations
  • Potential reflux from adjacent zones

Pressure changes the delivery problem

A positively pressurized enclosure can resist fog introduced from outside. A high free-air output rating does not guarantee entry through a long hose, small adapter or pressure boundary. Qualify the exact entry point, hose, adapter and system state before the formal study.

Use output conservatively

Inside a small enclosure, excessive fog can erase local detail and create condensation concerns. Start with controlled output and a delivery geometry that introduces the tracer upstream of the observation question without creating an artificial jet.

Applied Physics technical review

Build a defensible fogger requirement

Send Applied Physics the room, process, utilities, documentation and safety constraints—not just a model number.